Tuesday, June 15, 2010

Paula /FINSA and FORMALDEHYDE

When talking to Paula Health and Safety Finsa- she took issue with the fact that I repeatedly mentioned FORMALDEHYDE EMISSIONS-

I asked her if indeed FORMALDEHYDE is used in making chipboard - she said yes but wanted to know what particular agency had told me that emissions were Formaldehyde...
I told her that particular agency would be ME-
and RESEARCH

Formaldehyde Emissions

The manufacturing process to produce chipboard commonly uses Formaldehyde. Formaldehyde Emission is a toxic, carcinogenic gas, which affects the mucus membranes such as the lungs, eyes and nose. At low concentrations, Formaldehyde Emission effects cause a mild irritation, while at higher concentrations inhalation of Formaldehyde Emission can result in death. Egger (UK), a subsidiary of an Austrian company that has manufacturing sites in several European cities, manufactures chipboard in Northern England. In order for Egger (UK) to produce chipboard at this site, the HMIP (The UK pollution inspectorate) required that the plant measured, on a continuous basis, the vent emissions of formaldehyde to the atmosphere.

A Need for Multi-point Sampling of Formaldehyde Emissions

Traditional methods to measure Formaldehyde Emissions are normally time consuming, labour intensive and require expensive chemicals and laboratory facilities increasing the running costs. Egger (UK) decided that in addition to measuring vent emissions, they would also measure fugitive emissions and ambient levels within the factory boundary. This meant they required a multi-point sampling system that could simultaneously measure in 4 separate areas. They also wanted to be able to measure TOC levels during the process, which meant a multi-gas measurement capability must be included in the system. As legislation required vent emissions to be measured, the system must provide some form of hard copy of the measurement results as documentation to prove to the authorities that the process was running as planned. Egger (UK) also decided, as an extra safeguard for the workers, the system must provide real-time results that could be linked to an alarm system. This alarm system would be triggered if the concentrations of Formaldehyde Emission exceeded the permitted values.

The Tested Technique of Measuring Formaldehyde Emission

A Multi-gas INNOVA 1302 (superceded by the INNOVA 1412) was installed. This monitor, with it's photoacoustic (PAS) measurement technique and narrow band optical filters, is capable of measuring up to five gases simultaneously. The filters chosen for this application and the gases they measure are shown in Table 1 below.

GasesFilter NumbersDetection Limits
FormaldehydeUA09860.04ppm
MethanolUA09740.07ppm
DimethyletherUA09710.07ppm
TOC (ref. Toluene)UA09870.04ppm
Table 1: Gases to be measured and filters installed

The Solution

Although the UA0986 filter was chosen for Formaldehyde Emission, both Methanol and Dimethylether, present onsite, absorb in this filter's absorption range. Additional filters were included to enable cross-compensation for Methanol and Dimethylether. In addition to theINNOVA 1302, a sampler was installed. These were housed in an analyzer house within the plant. The vent was approx. 100m from the proposed instrument position (away from any zoned areas). The other sampling points were 50, 25 and 100m from the monitor. The high dew point in the scrubber meant that heating of the sampling system was needed, to avoid condensation in cold weather in the sample tube. Heated sampling lines were run through the plant to the sampling points. A coalescing filter was also installed as an additional safeguard to protect the monitor's measurement cell. An external pump was connected to the sampling lines enabling all the lines to be constantly purged to a common manifold. This reduced any measurement time delay (measurement results available every two minutes). Only the line to be sampled is diverted to the monitor. The monitor's own internal pump provides the sample. Result data for all four gases is passed, via an RS232 interface, to a printer to produce a hard copy.

The samples can be identified by the "channel number data" provided by the sampling system. The system has no expendable parts and the monitor's inherent stability means it only requires recalibrating approximately twice a year. So, once the measurement procedure is started the monitor can measure without any user intervention. The need for an alarm system is covered by the monitor's built-in beeper and Alarm Relay socket. The beeper warns staff working close to the monitor that the measured concentration has exceeded the alarm level, while the relay socket enables remote alarms to be connected to warn staff working in other parts of the factory. This is a feature Egger (UK) has earmarked for future development.

Conclusion

Even though the authorities demand a continuous monitoring system, which are often time consuming for those who have to operate them, by installing this system all the data and documentation necessary to satisfy the authorities are available without requiring more time from the workers. In addition to this, this monitoring system ensures the health and safety of the workers.

 Photoacoustic Field Gas Monitor 1412
The Photoacoustic Field Gas-Monitor 1412
Download Application Note
There are 1 files.








No comments:

Post a Comment